BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND PEN NDL 32G 4MM PRO 100 BOX 1200 US; Needle, hypodermic, single lumen
|
Back to Search Results |
|
Model Number 320550 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Pen needle will not attach to the insulin pen.
|
|
Manufacturer Narrative
|
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
|
Search Alerts/Recalls
|
|
|