Model Number DTMA1QQ
Device Problems Failure to Charge (1085); Delayed Charge Time (2586); Data Problem (3196)
Patient Problem Tachycardia (2095)
Event Date 01/24/2024
Event Type  Injury  
Event Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited a long charge time and a charge circuit timeout after reaching end of service (eos), resulting in aborted therapies for the patient's ventricular tachycardia (vt) episodes.It was also noted that the device displayed a false observation related to the feature which monitors the pacing amplitude threshold and adjusts lead outputs.Therapies were programmed off as the patient was entering hospice care.The device remains in the patient.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18703000
MDR Text Key335358379
Report Number9614453-2024-00629
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2019
Device Model NumberDTMA1QQ
Device Catalogue NumberDTMA1QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/24/2024
Date Device Manufactured11/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
429888 LEAD, 6935M55 LEAD, 5076-45 LEAD
Patient Outcome(s) Life Threatening;
Patient Age91 YR
Patient SexMale
Patient Weight79 KG
-
-