As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The analysis is thus based on the inspection of the manufacturing documents of the device as well as on the provided data.The quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The analysis of the provided trend data, acquired between (b)(6) 2024 and (b)(6) 2024 recorded an initial pacing impedance of 500 ohms with a drop to 300 ohms shortly after and an increase to 700 ohms at the end of recording period.Furthermore, an increasing shock impedance from 55 ohms to 75 ohms was recorded.The analysis of the provided iegm episodes revealed electromagnetic interference artifacts on the atrial, rv and far-field channel and the presence of the far-field on the ra channel, which led to oversensing.The provided x-ray images were analyzed.The perforation of the lead into the pericardium can be confirmed.In spite of the available information, no conclusion can be drawn regarding the root cause of the clinical observation.An analysis of the lead itself would be necessary to determine the root cause.Should additional information or the device itself become available for analysis, the investigation will be updated.
|