MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: returned product consisted of a jetstream xc-2.4 atherectomy catheter.The device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed no damage.There was blood in the device.Functional analysis was done by completing the setup procedure per the instructions for use (ifu).The device rotated as designed.Aspiration testing of the device was completed per the test procedure.The device is tested by using a 100ml beaker of water.The devices tip is submerged in a beaker of fluid and the device is run for a period of 1 minute.The final milliliters of fluid that the beaker shows after the 1minute run time is subtracted from the starting milliliters and the final number is the total that was aspirated.Test results showed that this device did not perform as designed per the test procedure specification sheet, withdrawing 26ml of fluid in the 1minute time frame.Inspection of the remainder of the device revealed no damage or irregularities.
 
Event Description
Reportable based on analysis completed on 25jan2024.It was reported that the device unexpectedly shutdown.A 2.4mm jetstream xc atherectomy catheter was selected for an atherectomy procedure in the superficial femoral artery.After the jetstream was primed, however, air bubbles continued to be seen in the tube during advancement to the lesion.The product was removed off the wire and reprimed.Once in the patient again, the jetstream unexpectedly shutdown and lost activity.The device was exchanged with a new jetstream catheter to complete the procedure, and no patient complications were reported.However, device analysis revealed only failure to evacuate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18707356
MDR Text Key335829560
Report Number2124215-2024-03916
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0032671477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-