ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY¿ CORE DR, MRI (US); implantable pacemaker Pulse-generator
|
Back to Search Results |
|
Model Number PM2140 |
Device Problems
Failure to Interrogate (1332); No Pacing (3268)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Further information was requested, but not received.
|
|
Event Description
|
Related manufacturer reference number: (b)(4), related manufacturer reference number: (b)(4).It was reported, that the patient presented to the clinic for a scheduled follow up.Upon interrogation of the pacemaker, the right ventricular (rv) and right atrial (ra) leads were found to have exhibited failure to capture.The physician elected for the rv and ra leads to be replaced.Prior to the procedure, the pacemaker could not be interrogated.And had no low voltage output, during the procedure.The pacemaker, rv and ra leads were all explanted and replaced.The patient was in stable condition throughout.
|
|
Manufacturer Narrative
|
The reported events of no output and inability to interrogate were confirmed.As received, the device had no telemetry communication and no output.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.The device was cut open to enable further testing and battery was found depleted.Hybrid circuitry was tested, resulting in high current drain, consistent with moisture damage, depleting the battery and resulting in the reported events.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.
|
|
Search Alerts/Recalls
|
|
|