BOSTON SCIENTIFIC CORPORATION ACCOLADE DR; IMPLANTABLE PACEMAKER Back to Search Results
Model Number L301
Device Problems Failure to Capture (1081); Over-Sensing (1438); Under-Sensing (1661); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Event Description
It was reported that this system with a pacemaker and right atrial (ra) lead showed loss of capture (loc) at maximum outputs in both bipolar and unipolar configurations.Furthermore, the lead showed intermittent under sensing of p-waves and intermittent over-sensing of the v on the a channel.The patient is not dependent.A chest x-ray was performed, and the mode was changed to ventricular pacing and sensing.The patient has an appointment with physician to evaluate changes and see if further action needs to be taken.At this time the ra lead and the pacemaker remain in service.No adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE DR
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18708646
MDR Text Key335408207
Report Number2124215-2024-08925
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970003/S167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/12/2018
Device Model NumberL301
Device Catalogue NumberL301
Device Lot Number747432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-