STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-209
Device Problems Material Fragmentation (1261); Naturally Worn (2988)
Patient Problem Joint Laxity (4526)
Event Date 01/23/2024
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.H3 other text : not returned.
 
Event Description
Patient had received a primary tka in 2012 though underwent revision surgery due to ongoing instability issues.Initial examination during surgery ((b)(6 2024) demonstrated increased medial/lateral laxity.Surgeon discovered fragments of initial cr size #2 9mm poly around the knee joint with significant wear on the posteromedial and posterolateral surfaces.Surgeon removed fragments of insert and replaced with a cs size #2 13mm poly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18710375
MDR Text Key335462155
Report Number0002249697-2024-00233
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K172634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Catalogue Number5530-G-209
Device Lot NumberLCQ915
Was Device Available for Evaluation? No
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-