Catalog Number 08057923190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was acceptable.The field service engineer (fse) determined that the analyzer is operating properly.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable igm-2 tina-quant igm gen.2 results for 1 patient sample on a cobas pro c 503 analytical unit.The initial igm result was 43 mg/dl.The customer was having issues with proficiency samples which prompted them to rerun qc and calibration, as well as perform a patient sample comparison.The sample was repeated the next day and the result was 31.3 mg/dl.
|
|
Manufacturer Narrative
|
Calibration was last performed on (b)(6) 2024.The alarm trace showed sample foam, sample clot, sample probe, and abnormal aspiration errors.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
|
Manufacturer Narrative
|
The medical coding for this case was updated.
|
|
Search Alerts/Recalls
|