Model Number 3851
Device Problems Failure to Advance (2524); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter facility name: (b)(6).E1: initial reporter address 1: (b)(6).
 
Event Description
It was reported that the tip was fractured.The 95% stenosed target lesion was located in the moderately tortuous and moderately left anterior descending artery (lad).A 10 mm x 3.50 mm wolverine coronary cutting balloon was selected for percutaneous coronary intervention (pci).During the procedure, the catheter could not cross lesion and it was noted that the tip of the device was fractured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable post procedure.
 
Manufacturer Narrative
E1: (b)(6).
 
Event Description
It was reported that the tip was fractured.The 95% stenosed target lesion was located in the moderately tortuous and moderately left anterior descending artery (lad).A 10 mm x 3.50 mm wolverine coronary cutting balloon was selected for percutaneous coronary intervention (pci).During the procedure, the catheter could not cross lesion and it was noted that the tip of the device was fractured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable post procedure.It was further reported that the fractured tip has been removed from the patient's body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18715131
MDR Text Key336197515
Report Number2124215-2024-08411
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0031364961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
Patient Weight88 KG
-
-