BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G138
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a request was made to have data from this cardiac resynchronization therapy defibrillator (crt-d) device analyzed.Data analysis confirmed that the power consumption is increasing in a gradual fashion.Device replacement was recommended.No adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a request was made to have data from this cardiac resynchronization therapy defibrillator (crt-d) device analyzed.Data analysis confirmed that the power consumption is increasing in a gradual fashion.Device replacement was recommended.No adverse patient effects were reported.Additional information received indicated that this crt-d device was explanted due to premature battery depletion (pbd).This device was replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18716224
MDR Text Key336507101
Report Number2124215-2024-09121
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public00802526589188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/07/2019
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number377365
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-