It was reported that this left ventricular (lv) lead exhibited high out range pacing impedance measurements of greater than 3000ohms.The field representative requested technical services (ts) to review the device electrogram (egm).Review of the presenting egm found high rates of lv sense likely due to noise.The lv lead was configured lv tip to right ventricular (rv) and the shock lead impedances appeared to be normal.Ts suspected that the cause of the high out of range impedances to be a lv tip issue and provided programming options.No adverse patient effects were reported.At this time, the lead remains in service.Subsequent additional information was provided by the field representative that based on the lead diagnostic findings, the physician suspected the lv lead to have been fractured.The lv lead was reprogrammed, and the issue was resolved.No additional adverse patient effects were reported.The lv lead remains in service.
|