SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.D5.Other operator of device: operator of device is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the motor of the device was making a loud noise.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found the following: enclosure was cracked near the quick connect down to the bottom, faded line cord, and the reservoir was leaking.The device also had outdated drain fitting, printed circuit board (pcb) and power switch.Functional testing confirmed the complaint.There was a loud noise.The root cause was attributed to an "old loud pump".The service history review identified there was no indication that the complaint was related to a service of the device within the review period.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18717541
MDR Text Key335721076
Report Number2183161-2024-00144
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/02/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-