The build lhr was examined including the final inspection records and the in-process measurements.There were no relevant ncmrs associated with this lot.The devices met the m6-c product specification including the axial stiffness and flexural resistance specifications.Rmf 0003 rev 38 line 12.75 - device explanted.Based on the inspection of the retrieved m6-c, the device did not show evidence of mechanical failure.Some in vivo damage was observed, including fraying of the fiber and a partial loss of furrows and abrasive damage to the core, however, inspection of all components, indicates that the device was likely intact at the time of removal.However, the lack of clinical data, radiographic analysis, and microbiological and histopathological analysis inhibits the further interpretation of the mechanism of failure of this device in this patient.
|