Model Number CARDIOHELP-I |
Device Problem
Device Alarm System (1012)
|
Patient Problem
Low Oxygen Saturation (2477)
|
Event Date 02/14/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
A getinge service technician will investigate the affected cardiohelp.A follow-up medwatch will be submitted when additional information becomes available.
|
|
Event Description
|
It was reported that a bubble alarm tripped and could not be cleared out.The failure occurred during treatment.There was no indication of a bubble.The emergency drive was used until the cardiohelp was exchanged.The patient de-saturated into the 20% range.Complaint id# (b)(4).
|
|
Manufacturer Narrative
|
A getinge field service technician (fst) was sent for investigation on 2024-03-15.The cardiohelp function was tested for 15 days at 3200rpm and a flow of 4.8 lpm.No failure occurred.Therefore no part was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.A follow-up medwatch will be submitted when additional information becomes available.
|
|
Event Description
|
Complaint id# (b)(4).
|
|