Model Number 865351 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Reporting address state: (b)(6).
|
|
Event Description
|
The customer reported that the sound is very low or cannot be heard.The device was in use on a patient at the time of the event.There was no adverse event reported.
|
|
Manufacturer Narrative
|
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that no speaker sound at start up test because the speaker was defective.The cause of the reported problem was a defective speaker.The reported problem was confirmed.The customer was provided a replacement device to resolve the issue.
|
|