Catalog Number 03003523001 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The serial number of the integra 400 plus analyzer is (b)(6).The field service engineer determined there was a poorly fitting sealing ring on the reference electrode, causing air to enter.The electrode had recently been replaced.The sealing rings were cleaned and checked for integrity; they were in acceptable condition.The cl electrode was changed.Diagnostic tests were acceptable.Maintenance was performed on the ise module.Calibration and controls were run and expected values were measured.The investigation is ongoing.
|
|
Event Description
|
The initial reporter stated they received discrepant results for one patient sample tested with the chloride electrode on a cobas integra 400 plus.No questionable results were reported outside of the laboratory.The sample resulted in a cl value of 120 mmol/l when tested on the integra 400 plus analyzer.The sample was repeated on an unknown gasometer instrument, resulting in a cl value of 109 mmol/l.This value was considered to be correct by the customer.
|
|
Manufacturer Narrative
|
The investigation determined the service actions resolved the issue.
|
|
Search Alerts/Recalls
|