SYNTHES GMBH PROTECT SLEEVE 16/11 F/PFNA BLADE; GUIDE Back to Search Results
Catalog Number 356.818
Device Problems Device-Device Incompatibility (2919); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Event Description
Device report from australia reports an event as follows: it was reported that on 07-february-2024 during loan kit inspection, the products in question were found to have damage.The threads of the protection sleeve were burred, causing the compression nut to become stuck.There was no reported patient or procedure interaction.No further information is available.This report is for a protect sleeve 16/11 f/pfna blade.This is report 1 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a synthes employee.G4: device is not distributed in the united states, but is similar to device marketed in the usa.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that 356.818, protect sleeve 16/11 f/pfna blade had nicks on the threads and device appears to be assembled.Based on damage occurred we can confirm the unable to assemble/disassemble reported event.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the 356.818, protect sleeve 16/11 f/pfna blade would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device product code: 356.818, lot number: 880p746, it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 27-jul-2022, manufacturing site:jabil bettlach.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECT SLEEVE 16/11 F/PFNA BLADE
Type of Device
GUIDE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
JABIL BETTLACH
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18721265
MDR Text Key335583213
Report Number8030965-2024-02583
Device Sequence Number1
Product Code FZX
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number356.818
Device Lot Number880P746
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
BUTTRESS/COMPR-NUT F/PFNA BLADE
-
-