Model Number RONYX40012X |
Device Problems
Difficult to Remove (1528); Activation, Positioning or Separation Problem (2906)
|
Patient Problems
Device Embedded In Tissue or Plaque (3165); Insufficient Information (4580)
|
Event Date 01/31/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
An attempt was made to use one resolute onyx coronary drug-eluting stent to treat a lesion in the ostium of the left main (lm) coronary artery.The device was not inspected.It was reported that there was difficulty removing the device following stent deployment.It was detailed that when the stent was check, it was found that the stent frame did not deploy and the vessel did not expand.Another stent was used without removing the resolute onyx stent.The patient is alive with no further injury.
|
|
Manufacturer Narrative
|
Additional information: an attempt was made to use one resolute onyx coronary drug-eluting stent to treat a lesion in the ostium of the left main (lm) coronary artery.There were no issues noted when removing the device from the hoop/tray.Resistance was noted while advancing the device to the lesion.Excessive force was not used.The stent could not be removed from the vessel when it was noted that the stent did not deploy.Another stent (unknown brand) was used to fully deploy the ronyx without removing this stent from the patient.Patient's gender added.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|