• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation Back to Search Results
Model Number A-TFSE-DF
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Event Description
This report is to advise of a displaced electrode and fracture in the conductor wire.
 
Manufacturer Narrative
One bi-directional, curve d-f, tactiflex ablation catheter, sensor enabled was received for evaluation.Electrode 2 read as an open circuit during electrical testing, consistent with the reported event.Further investigation revealed electrode ring 2 was shifted distally on the catheter shaft causing a fracture in conductor wire 2, consistent with the open circuit detected.The shaft outer diameter at electrode ring 2 was not within specifications, however this was determined to not have contributed to the shifted electrode ring or the fractured wire.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the shifted electrode ring and subsequent fracture remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿
Type of Device
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18721370
MDR Text Key335584119
Report Number3008452825-2024-00092
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P220013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model NumberA-TFSE-DF
Device Lot Number8955663
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
Patient SexFemale
Patient Weight52 KG
-
-