BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (TPE); SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA Back to Search Results
Catalog Number 955470
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2024
Event Type  malfunction  
Event Description
It was reported that an external fluid leak was observed from the waste bag of a prismaflex tpe 2000 set during continuous renal replacement therapy.The machine did not alarm.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, photographs and a video were provided for evaluation.During visual inspection of the provided video, a leak was observed from the drain bag sealing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was determined to be related to misuse of the bag.The drain bag is a single use product and must not be emptied and reused during the same therapy.A previous nonconformance investigation concluded that the filling/emptying cycles during the same therapy led to physical constraints on the bag, causing pinholes to form that allowed leakage.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (TPE)
Type of Device
SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18723026
MDR Text Key336542870
Report Number8010182-2024-00057
Device Sequence Number1
Product Code MDP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955470
Device Lot Number22B0071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-