The actual device was not available; however, photographs and a video were provided for evaluation.During visual inspection of the provided video, a leak was observed from the drain bag sealing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was determined to be related to misuse of the bag.The drain bag is a single use product and must not be emptied and reused during the same therapy.A previous nonconformance investigation concluded that the filling/emptying cycles during the same therapy led to physical constraints on the bag, causing pinholes to form that allowed leakage.Should additional relevant information become available, a supplemental report will be submitted.
|