Investigation conclusion the investigation of the returned coil system found the pusher returned without the implant attached, and the pusher hypotube kinked at the proximal section.The implant was returned separated from the pusher with the overcoil stretched and was stuck within the middle section of the microcatheter.The unit did not pass continuity and resistance testing.Further inspection of the pusher found the green lead wire broken and the clear pet damaged at the proximal solder joint, which is consistent with non-detachment.The pusher's monofilament showed a tensile break shape at the tip, which is consistent with the device experiencing force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.The microcatheter used in the procedure was found damaged at multiple areas, which may have caused or contributed to the separation of the coil from the pusher.
|