Catalog Number 04906969001
Device Problem Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2024
Event Type  malfunction  
Manufacturer Narrative
The test strip lot number and expiration date were requested, but not provided.The investigation is ongoing.
 
Event Description
The initial reporter stated they received questionable results for an unspecified number of patient samples tested on a cobas u 411 urine analyzer.Multiple test parameters are affected.The customer provided an example of one patient sample with discrepant leukocyte values.The sample initially resulted in a leukocyte value of 500 leukocytes/ul.The same sample was repeated, resulting in a negative leukocyte value.
 
Manufacturer Narrative
The quality controls failed.The field service engineer replaced the photometer and the instrument was working properly after this action.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U 411 URINE ANALYZER
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18725239
MDR Text Key335611680
Report Number1823260-2024-00467
Device Sequence Number1
Product Code KQO
Combination Product (y/n)Y
Reporter Country CodePL
PMA/PMN Number
K093555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04906969001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-