SMITHS MEDICAL ASD, INC. CADD-LEGACY PCA AMBULATORY INFUSION PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 6300
Device Problems Device Alarm System (1012); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
H3: device not received by manufacturer.D4 - serial #: reported as either (b)(6) or (b)(6).H4: unknown.No product was returned.The investigation determined the most probable cause to be a main board failure, however this cannot be confirmed as no product was returned for investigation.If the product is returned this complaint will be reopened for further investigation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Event Description
It was reported that during use, the patient presented to the emergency department because the pump would not stay running; it kept alarming and going into stop mode.The batteries were changed but the pump continued to alarm every 10 minutes.The pump was replaced.Per the reporter, there were no adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY PCA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18727403
MDR Text Key336638422
Report Number3012307300-2024-00863
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982839
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number6300
Device Catalogue Number21-6300-51
Was Device Available for Evaluation? No
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-