Model Number 6300 |
Device Problems
Device Alarm System (1012); Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3: device not received by manufacturer.D4 - serial #: reported as either (b)(6) or (b)(6).H4: unknown.No product was returned.The investigation determined the most probable cause to be a main board failure, however this cannot be confirmed as no product was returned for investigation.If the product is returned this complaint will be reopened for further investigation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.
|
|
Event Description
|
It was reported that during use, the patient presented to the emergency department because the pump would not stay running; it kept alarming and going into stop mode.The batteries were changed but the pump continued to alarm every 10 minutes.The pump was replaced.Per the reporter, there were no adverse patient effects.
|
|