Device evaluated by mfr.Synergy xd mr ous 3.50 x 12mm stent delivery system (sds); catheter number was returned for analysis.A visual and tactile examination of the hypotube found multiple kinks along the shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No stent was returned for analysis.The balloon cones were reviewed, and the balloon was in a deflated state.Based on microscopic analysis, no stent was returned for analysis.A review of the manufacturing stent profile data was performed and the stent od at the time of manufactured.The balloon cones were reviewed, and the balloon was in a deflated state.A visual and tactile examination of the hypotube found multiple kinks along the shaft.Bumper tip showed no signs of distal tip damage.
|