|
Model Number PM3262 |
Device Problems
Failure to Interrogate (1332); No Pacing (3268)
|
Patient Problem
Arrhythmia (1721)
|
Event Date 01/28/2024 |
Event Type
Injury
|
Event Description
|
It was reported, that the patient presented in emergency room, due to an arrhythmia.Device interrogation revealed, failure to capture, failure to interrogate with no magnet and no radiofrequency.And no low voltage output on the pacemaker.The physician elected to explant and replace the pacemaker.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported events of no magnet response, no rf telemetry and no output were confirmed.As received, the device had no telemetry communication and no output.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.The device was cut open to enable further testing and battery was found depleted.Hybrid circuitry was tested, resulting in high current drain, consistent with moisture damage, depleting the battery and resulting in the reported events.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.
|
|
Event Description
|
Correction: it was reported that the patient presented in emergency room due to a bradycardic arrhythmia.Device interrogation revealed failure to interrogate with both radiofrequency and inductive telemetry, no magnet response, and no low voltage output on the pacemaker.The physician elected to explant and replace the pacemaker.The patient was in stable condition.
|
|
Search Alerts/Recalls
|
|
|