• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the mayfield skull clamp (a1059) was loose and would not tighten on the patient's head.No injury or surgical delay has been reported.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: device history review (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit showed that the lock had both rotational and lateral movement, and a residue buildup was present.The index knob and the lock needed new components due to worn internal parts; unit will need to be machined to have heli-coils added to large starburst threads.General cleaning and maintenance will be performed.Root cause - the complaint is confirmed via inspection of the unit.The index knob and lock required replacement of worn internal components due to routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18727976
MDR Text Key336321231
Report Number3004608878-2024-00020
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2024
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-