• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 71675383
Device Problem Fracture (1260)
Patient Problem Failure of Implant (1924)
Event Date 01/16/2024
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that, after a trauma surgery was performed on (b)(6) 2023, the patient fell and broke said implant.This adverse event was treated by a revision surgery on (b)(6) 2024, in which a intertan 10s 10mm x 18cm 125d was exchanged to correct the broken implant.Patient's current health status is unknown.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, clinical/ medical records have not been provided; however, it¿s noted the patient fell and broke the implant.The patient impact is a revision and the expected post-operative convalescence period.No further clinical assessment is warranted at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for intramedullary nail system revealed in possible adverse effects that cracking or fracture of the implant may occur.Preoperative planning section states that proper type and size of implant must be selected to ensure effective treatment of patients considering factors such as patient¿s size, strength, skeletal characteristics, skeletal health, and general health.Overweight or musculoskeletal deficient or unhealthy patients may create greater loads on implants that may lead to breakage or other failure of the implants.Additionally, postoperative care section warns that early weight bearing substantially increases implant loading and increases the risk of breaking the device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include size selected, surgical technique, postoperative care, patient anatomy, abnormal loading of limb, excessive forces and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 18CM 125D
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18728603
MDR Text Key335705885
Report Number1020279-2024-00386
Device Sequence Number1
Product Code JDS
UDI-Device Identifier00885556105696
UDI-Public00885556105696
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040212
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number71675383
Device Lot Number22KM07266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2024
Date Device Manufactured10/11/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-