MAUDE Adverse Event Report: LENSAR, INC ALLY ADAPTIVE CATRACT TREATMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSAR, INC ALLY ADAPTIVE CATRACT TREATMENT SYSTEM Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Capsular Bag Tear (2639)
Event Date 01/16/2024
Event Type  Injury  
Manufacturer Narrative
The suction ring in the procedure is locked properly.Laser completed capsulotomy with little to no patient movement.During fragmentation a large gas bubble forms behind the lens.There is no indication the laser did not function as designed.Anterior vitrectomy performed and 3-piece lens placed.
 
Event Description
P1008 - posterior capsular rupture issue: on (b)(6) 2023, ginger powell ((b)(6)) reported to jessica byer that dr.(b)(6) had a posterior rupture on case #(b)(4).Anterior vitrectomy performed and 3-piece lens placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLY ADAPTIVE CATRACT TREATMENT SYSTEM
Type of Device
ALLY ADAPTIVE CATRACT TREATMENT SYSTEM
Manufacturer (Section D)
LENSAR, INC
2800 discovery drive ste.100
orlando FL 32826
Manufacturer (Section G)
LENSAR, INC
2800 discovery drive ste. 100
orlando FL 32826
MDR Report Key18730487
MDR Text Key335764157
Report Number3009026057-2024-00006
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182795
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number70-00050-001
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-