It was reported that after successfully implant completion of right atrial (ra) lead, the lead became dislodged and noticed before the pocket was closed.Upon repositioning the lead, unable to fully advance multiple stylets.The lead was replaced with no patient consequences.
|
A complete lead was returned in one piece.The reported event of difficult to insert stylet was confirmed.Visual inspection of the lead found the helix to be retracted and clogged with blood.X-ray examination found the inner coil to be over torqued at the connector region, consistent with procedure damage.After cutting, cleaning, and applying torque directly to the inner coil, the helix was able to extend and retract.The measured full helix extension length was within product specification.Electrical testing did not find any indication of conductor fractures or internal shorts.The cause of the reported event difficult/unable to insert the stylet was due to an over torqued inner coil at the connector region, consistent with procedure damage.
|