MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM G; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Joint Dislocation (2374)
Event Date 03/07/2023
Event Type  Injury  
Event Description
It was reported that the patient was revised approximately two years post-implantation due to a dislocation.The head was revised.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
Cmp-(b)(4).D10: cat# 00877703603 lot# 2963426 bioloxâ® option, head, l, 㸠36/+3.5, taper 12/14 g2: foreign ¿ new zealand.The customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were reviewed and identified the following: the patient underwent an initial left tha.The patient was revised on one year later due to dislocation and the head was exchanged.A second revision occurred another year later due to a second dislocation, where the head and liner were exchanged.A definitive root cause cannot be determined.This complaint was confirmed based on the provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 36MM G
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18731593
MDR Text Key335719721
Report Number0001822565-2024-00592
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010000937
Device Lot Number6854147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-