|
Model Number 83779 |
Device Problems
Entrapment of Device (1212); Stretched (1601); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that part of the coil protruded from catheter tip upon removal.The target lesion was in the arteriovenous fistula.A 8mm x 20cm pe f-idc interlock was selected for use.During the procedure, it was noted that the coil was pushed into the microcatheter, a great resistance was felt.The coil could not be pushed out after being pushed halfway.After repeated attempts, it still could not be pushed out into the distal of the catheter.The device was removed from the patient together with the catheter with part of the coil protruding from the catheter tip.It was also found that the coil had been stretched upon removal.The procedure was completed with another of the same device.No complications reported and the patient was in stable condition after procedure.
|
|
Event Description
|
It was reported that part of the coil protruded from catheter tip upon removal.The target lesion was in the arteriovenous fistula.A 8mm x 20cm pe f-idc interlock was selected for use.During the procedure, it was noted that when the coil was pushed into the microcatheter, a great resistance was felt.The coil could not be pushed out after being pushed halfway.After repeated attempts, it still could not be pushed out into the distal end of the catheter.The device was removed from the patient together with the catheter with part of the coil protruding from the catheter tip.It was also found that the coil had been stretched upon removal.The procedure was completed with another of the same device.No complications were reported and the patient was in stable condition after procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.Visual inspection revealed that only the main coil was returned.It was observed that the main coil was bent and stretched at the primary and zap tip section.Functional testing could not be performed since only the main coil was returned.The coil arm outer diameter (od), zap tip od, and primary coil od passed the dimensional inspection.
|
|
Search Alerts/Recalls
|
|
|