WILSON-COOK MEDICAL INC HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC; KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Catalog Number HBD-W-10-11-12
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During a colon balloon dilation, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.It was reported [that] upon inflating balloon, the balloon would not inflate properly and when they were pulling the balloon back out of the scope it was noted that the catheter was broken (snapped), which was causing the balloon to not inflate properly.Another balloon was used to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the balloon fully deflated and the wire still inside the device.The catheter was cracked/broken at 116.3cm distal from the handle.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report, there was a break in the catheter.A corrective action (capa) has been initiated to reduce occurrences of catheter cracking, splitting or breaking for hbd-w devices.The product said to be involved is included in the scope of the corrective actions.In the report it does state it was unknown if negative pressure was applied to the balloon prior to advancing down the endoscope accessory channel.The ifu states: "to facilitate passage through the endoscope, apply negative pressure to the device." "maintain balloon deflation with negative pressure and introduce into endoscope accessory channel, advancing in short increments until balloon is completely visualized endoscopically." prior to distribution, all hercules 3 stage wire guided balloons esophageal-pyloric-colonic are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC
Type of Device
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key18735512
MDR Text Key336740274
Report Number1037905-2024-00092
Device Sequence Number1
Product Code KNQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHBD-W-10-11-12
Device Lot NumberW4775529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2024
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ENDOSCOPE-OLYMPUS PCF-190
-
-