Model Number IPN918765
Device Problems Difficult to Remove (1528); Material Separation (1562)
Patient Problem Insufficient Information (4580)
Event Date 01/24/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported "swg removed another kit was used (with a similar incident occurring), it was not the same insertion site.No medical intervention required.The patient was taken for ct angiography for vascular axes.The patient's current condition is reported as "critical".See associated mdr number 3006425876-2024-00119.
 
Event Description
It was reported "swg removed another kit was used (with a similar incident occurring), it was not the same insertion site.No medical intervention required.The patient was taken for ct angiography for vascular axes.The patient's current condition is reported as "critical".See associated mdr number 3006425876-2024-00119.
 
Manufacturer Narrative
Qn#( b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.Warning: do not withdraw guidewire against needle bevel to reduce risk of possible severing or damaging of guidewire." a device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CVC SET: 3-LUMEN 7 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18737961
MDR Text Key336880563
Report Number3006425876-2024-00131
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN918765
Device Catalogue NumberCV-15703
Device Lot Number71F23D3073
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NONE REPORTED.; NONE REPORTED.
-
-