• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106015
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that the patient had an external driveline repair in (b)(6) 2023.They went into clinic recently to monitor the cable for phase c degradation.Left ventricular assist device was stable.Log files captured a few driveline fault events on (b)(6) 2024 at 18:43 and (b)(6) 2024 at 12:39 and 12:54.It appeared that the orange wire was having some intermittent continuity issues but was not broken as the events were very intermittent.The other 5 driveline wires appeared to be functioning as intended.
 
Manufacturer Narrative
Related event that covers the driveline repair on (b)(6) 2023 is reported under mfr#: 2916596-2023-05310.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS)
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18740522
MDR Text Key336749160
Report Number2916596-2024-00770
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexMale
Patient Weight86 KG
-
-