NIHON KOHDEN ORANGEMED, LLC NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number NKV-550-U
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2024
Event Type  malfunction  
Manufacturer Narrative
Nihon kohden orangemed will submit a supplemental report if additional information becomes available.
 
Event Description
It was reported by the registered nurse (rn) that the ventilator made a high-pitched alarm sound, and the graphic user interface touchscreen went black.When the respiratory therapist (rt) arrived at the patient''s bedside, the ventilator was in the process of restarting.After 30-60 seconds, the touchscreen displayed previously set settings and mode.After acknowledging the settings, the rt selected resume ventilation and the device functioned normally.The patient remained on this ventilator until the next night and then was placed on another ventilator.There was no harm or serious injury to the patient.Vital signs were unchanged.
 
Manufacturer Narrative
Evaluation completed.See attached failure investigation summary report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NIHON KOHDEN ORANGEMED, LLC
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer (Section G)
NIHON KOHEN ORANGEMED, LLC
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer Contact
sheryl higgins
1800 e. wilshire ave.
santa ana, CA 92705
949502644
MDR Report Key18740635
MDR Text Key336306338
Report Number3014631252-2024-00008
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192307
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberNKV-550-U
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age5 DA
Patient SexMale
Patient Weight2 KG
-
-