Model Number 102956
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Event Description
It was reported that during patient support the centrimag console issued an s3 alarm.The fan was checked for blockages or for overheating.The console was cleaned.Approximately 15 minutes later, the alert disappeared and the computer continued to function normally.There were no patient consequences.Console mfr # 2916596-2024-00803.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.Section e1: reporter postal office or zip code: bogotá section e1: reporter address line 1: cr 16 93 a 36 of 804.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18741591
MDR Text Key336774355
Report Number3003306248-2024-00425
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-