SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-390 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: date of event and d5.Operator of device is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during preventative maintenance, the device was found not alarming and no leds.Board needs to be replaced.No patient or clinical harm reported.
|
|
Manufacturer Narrative
|
H6.Investigation codes: updated.Investigation summary: the affected device was returned for evaluation.A visual inspection was performed, and the led lights were found to be broken.The customer's indicated failure was confirmed as the device alarmed and the led lights were damaged.The root cause was the damaged pcb.No action was taken due to the demand, condition, and age of the device.It was deemed beyond economical repair and will be scrapped.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.D4.Unique identifier (udi) #: (b)(4).
|
|