SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. CORRUGATED ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101330-NLJ |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3.Date of event: month and year of event have been provided, but day is unknown.D4.Lot number, udi, expiration date and h4.Manufacture date are unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the device failed a leak test.There was no patient involvement and no patient harm/adverse event reported.
|
|
Manufacturer Narrative
|
H3 and h6 - evaluation codes: updated.Device evaluation: one used unit and three photos were returned for investigation.During visual inspection, no defects were found.Functional testing found a leak was observed in the circuit, failure mode was confirmed.Root cause was not established.No lot number was provided or found therefore no device history report (dhr) could be performed.This issue will continue to be monitored and further actions taken accordingly.
|
|