Model Number 4003 |
Device Problem
Sensing Intermittently (1558)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the sensor provided solid values on the monitor and then the reading just flatlined for no apparent reason.There was no patient impact or consequence reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned sensor was evaluated.The sensor had no damage or defects during the visual inspection.The sensor passed continuity, and functional testing.During additional testing, there were no intermittent readings observed.The sensor was determined to be functioning as designed.
|
|
Event Description
|
The customer reported the sensor provided solid values on the monitor and then the reading just flatlined for no apparent reason.There was no patient impact or consequence reported.
|
|
Search Alerts/Recalls
|