SPACELABS HEALTHCARE XHIBIT CENTRAL STATION Back to Search Results
Model Number 96102
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Manufacturer Narrative
The customer stated that while telemetry monitoring was unavailable, clinical staff manually monitored staff.Tech support walked the user through restarting their xhibit central station.Following the restart, the patient zones were unassigned, tech support walked the user through reassigning patient zones to continue monitoring.While patient monitoring was resumed, cause of failure was not established.
 
Event Description
The customer reported that several telemetry beds went offline while monitoring patients.There was no patient or user harm associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XHIBIT CENTRAL STATION
Type of Device
XHIBIT CENTRAL
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key18744456
MDR Text Key335854141
Report Number3010157426-2024-00029
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number96102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-