• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC CATALYS SYSTEM; OPHTHALMIC FEMTOSECOND LASER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC CATALYS SYSTEM; OPHTHALMIC FEMTOSECOND LASER Back to Search Results
Model Number CATALYS-U
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Eye / Vision Problem (4471)
Event Date 01/19/2024
Event Type  Injury  
Event Description
It was reported that the surgery had poor wound closure from the primary incision and the surgeon used visco to seal the wound.No additional information was provided.
 
Manufacturer Narrative
Section a4 and a5: account did not have this information device evaluation: an application support manager (asm) was at the site and advised the doctor to change the anterior cut angle of the wound to help with closure.Asm updated the anterior side cut angle from 70 to 90.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATALYS SYSTEM
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key18744561
MDR Text Key335811054
Report Number3012236936-2024-00442
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113479
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCATALYS-U
Device Catalogue NumberCATALYS-U
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexMale
-
-