BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U228 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/07/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that the lead safety switch (lss) had been triggered for this system due to a pacing impedance measurement greater than 3000 ohms on the right ventricular channel.Additionally, myopotential noise and elevated threshold measurements were observed.The physician reprogrammed the device back to bipolar configuration and no noise was observed, and impedance measurements were within range.The physician left the device in bipolar sensing and unipolar pacing.A boston scientific technical services consultant documented the reported clinical observations.The system remains implanted.The associated rv lead is manufactured by a boston scientific competitor.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that the lead safety switch (lss) had been triggered for this system due to a pacing impedance measurement greater than 3000 ohms on the right ventricular (rv) channel.Additionally, myopotential noise and elevated threshold measurements were observed.The physician reprogrammed the device back to bipolar configuration and no noise was observed, and impedance measurements were within range.The physician left the device in bipolar sensing and unipolar pacing.A boston scientific technical services consultant documented the reported clinical observations.The system remains implanted.The associated rv lead is manufactured by a boston scientific competitor.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|