Catalog Number 010000664 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/30/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during the procedure, while removing the screw hole covers, one was not able to be removed from the shell.Several attempts were made before the surgeon decided to use a new shell to complete the procedure.There was no harm or health impact to the patient.Attempts have been made and no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4; b5; g3; h2; h3; h6.Upon visual inspection there is no damage to the shell however the hex feature of the plug has been stripped.Due to the stripping the plug could not be removed for further evaluation.The complaint is confirmed via evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|