• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZ9270
Device Problems Leak/Splash (1354); Separation Problem (4043)
Patient Problems Hypoglycemia (1912); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2024
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd maxzero multi-fuse extension set with needleless connector separated.The following information was received by the initial reporter with the following: date of occurrence: on (b)(6) 2024.Description of what happened: infant with low blood sugar reading <25.Ivf d10w infusing.Rn noted trifuse separated at the filter.Rn changed trifuse, blood sugar came up with bolus of d10w, product#: and lot#: mz9270, lot#: 22119145, was the course of treatment altered? neonate required bolus of d10w to bring blood sugar up.Any harm to the patient? mild temporary harm.
 
Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint of separation at tubing to filter connection could not be verified due to the product not being returned for failure investigation.Device history record review for model mz9270, lot number 22119145 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 19dec2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause of this failure could not be identified without a failure investigation.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18748335
MDR Text Key335934563
Report Number9616066-2024-00274
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140831
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMZ9270
Device Lot Number22119145
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-