Catalog Number TMMC9609L |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that a blood recipient set had hair in an unspecified location.This was observed before the set went to the customer.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
B5: additional information was received that the hair was inside the package "on top of the tubing".H4: the lot was produced on july 2023.H10: the actual device was not available; however, retained samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The retained samples were gravity then leak tested under water; no leaks were observed.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|