It was reported that this pacemaker system stored two signal artifact monitor (sam) episodes at the time of the patient's unrelated shoulder surgery.There was oversensed electromagnetic interference (emi) noise on all channels with less than two seconds of pacing inhibition, and the minute ventilation (mv) feature was disabled at that time.Additionally, review of a live electrogram (egm) revealed 4-5 beats of atrial tachycardia below the sensing floor, resulting in right atrial (ra) undersensing.Adjustments to ra sensitivity resolved the observed undersensing, and the mv feature was re-enabled.Technical services (ts) reviewed troubleshooting options and other programming considerations.This pacemaker system remains in service.No additional adverse patient effects were reported.
|