Model Number L111
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Failure to Read Input Signal (1581); Under-Sensing (1661); Use of Device Problem (1670)
Patient Problem Bradycardia (1751)
Event Date 02/05/2024
Event Type  malfunction  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this pacemaker system stored two signal artifact monitor (sam) episodes at the time of the patient's unrelated shoulder surgery.There was oversensed electromagnetic interference (emi) noise on all channels with less than two seconds of pacing inhibition, and the minute ventilation (mv) feature was disabled at that time.Additionally, review of a live electrogram (egm) revealed 4-5 beats of atrial tachycardia below the sensing floor, resulting in right atrial (ra) undersensing.Adjustments to ra sensitivity resolved the observed undersensing, and the mv feature was re-enabled.Technical services (ts) reviewed troubleshooting options and other programming considerations.This pacemaker system remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18749399
MDR Text Key336787785
Report Number2124215-2024-10370
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/11/2024
Device Model NumberL111
Device Catalogue NumberL111
Device Lot Number600802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-