Model Number V60
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 device indicating that error codes 1113 "barometer calibration data error", 110f "o2 flow sensor calibration data error", 1107 " proximal pressure sensor calibration data error" occurred.At this time, it is unknown if there was patient involvement at the time the issue was discovered; however, there was no reported harm to the patient or user.
 
Manufacturer Narrative
Error 1006 "da pcba adc failed" also occurred.There was no patient involvement at the time the error codes were discovered as the device was not in use and was pulled from service.A field service engineer (fse) was dispatched onsite to evaluate and repair the device.Upon arrival, the fse spoke with the biomedical engineer (bme) about the history of the device and inspected the device.The fse then let the device run and cycle normally and reviewed the information screen and event logs.The fse noticed that components were replaced recently prior his visit and discovered that the error codes in question were from august 2023.The fse was not able to recreate the error codes while testing and running the device.The error codes could not be duplicated.The fse confirmed that the device successfully passed performance verification and electrical safety testing and was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18751461
MDR Text Key336068258
Report Number2518422-2024-08945
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2024
Date Device Manufactured03/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-