• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BELLCO SRL CARPEDIEM SYSTEM; PEDIATRIC HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BELLCO SRL CARPEDIEM SYSTEM; PEDIATRIC HEMODIALYSIS SYSTEM Back to Search Results
Model Number IB0595540
Device Problems Device Sensing Problem (2917); Protective Measures Problem (3015)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 01/06/2023
Event Type  Injury  
Manufacturer Narrative
D10 concomitant products: ib0595540, cc ib0595540, kit cvvhd 015 carpediem, (lot#: 2201000320).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, a 24-day-old patient was hospitalized.For spontaneous prematurity with acute anuric renal failure requiring peritoneal dialysis.During the dialysis session, the membrane of the pressure sensor on the dialysis machine cracked after 30 minutes of dialysis on the circuit.The problem repeated with another dialysis kit at the time of connection.The alarm was activated.There was no issue with the machine.Nothing unusual was observed, on the kit prior to use.Priming was done, and the result was normal.They ended the treatment, changed the kit with the same product id (identifier) to resolve the issue.And the treatment was completed with the replacement kit.The clinical consequences and current state of the patient were that two dialysis were delayed.The patient was released.Disorganization and lengthening of the duration of care for around an hour.The patient had little blood loss.And the exact volume lost was unknown.A blood transfusion was the intervention or treatment required as a result of the event.And the exact value of the volume was unknown.The blood transfusion was, due to the small loss of blood.The patient was premature and had a low volume of blood.That was why, even if the blood loss was small, the hcp (heath care professional) had to replace it.The patient was in stable condition, but the anuric renal failure still had to be treated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPEDIEM SYSTEM
Type of Device
PEDIATRIC HEMODIALYSIS SYSTEM
Manufacturer (Section D)
BELLCO SRL
v camurana 1
mirandola,mo 41037
IT  41037
Manufacturer (Section G)
BELLCO SRL
v camurana 1
mirandola,mo 41037
IT   41037
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18751621
MDR Text Key335935795
Report Number1000312731-2024-00004
Device Sequence Number1
Product Code QIR
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
DEN180055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIB0595540
Device Catalogue NumberIB0595540
Device Lot Number2201000320
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2024
Date Device Manufactured01/26/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
-
-