HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1103 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Stroke/CVA (1770); Muscle Weakness (1967); Paralysis (1997); Thrombosis/Thrombus (4440)
|
Event Date 01/31/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the patient was hospitalized for a major stroke.The patient experienced significant deficits of paralysis, weakness, and speech deficits and would no longer be independent.It was noted that the patient's anticoagulation was therapeutic.Computerized tomography (ct) confirmed an ischemic stroke; it was suspected that the ventricular assist device (vad) threw a thrombus.It was reported that the patient's risk for another stroke was high.The patient had an improved left ventricular ejection fraction (lvef) of 40%, so the vad was decommissioned and the patient underwent inpatient rehab. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion. medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation completion.Product event summary: (b)(6) was not returned for evaluation.Log file analysis could not be performed since log files were not available for analysis.As a result, the reported thrombus event could not be confirmed due to insufficient information.Based on the available information, the device may have caused or contributed to the reported event.Per the instructions for use, neurological dysfunction and thrombus are known potential complications associated with the implantation of a vad.Based on review of past adverse events for this patient, it was noted that the patient had a history of neurological dysfunction.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|