• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2024
Event Type  malfunction  
Event Description
It was reported that the device was too loud.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.D5.Other operator of device: operator of device is unknown.Investigation summary: the affected device was received for evaluation and was visually inspected to determine that it was in good physical condition, with small amounts of wear and tear on the enclosure.The device was unpacked, the tank was filled with water and a disposable was attached.The start button was pressed and right away it was noticed that the device was too loud, confirming the customer's indicated failure.The device cover was removed, and the chassis was making contact with the encloser on the bottom.This was determined to be the root cause of the issue.The chassis was rearranged from the 4 screws for the pole clamp, and it was pushed a little bit higher.After arranging it, the device was not noisy anymore.After performing the entire repair, a functional test was performed, and after calibration the temp was stable and in range.The o-rings and tank gasket were also replaced.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18759605
MDR Text Key336179593
Report Number2183161-2024-00162
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90-INT-230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-